Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01020-1 |
Product Name/Description |
Optimys Opening Broach Bent (Reusable surgical instrument used for opening the medullary canal)
Article Number: 51.34.0079
Lot Number: 6054883
ARTG Number: 184413 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
17/10/2013 |
Responsible Entity |
|
Reason/Issue |
Mathys Ltd Bettlach has been notified of an incident in Switzerland where the optimys Opening Broach broke intraoperatively and the fragment was left in the bone. In the course of the investigations, it became apparent that the Opening Broach Bent may exhibit surface defects in the hewn portion which may favour breakage in case of non-compliant usage. |
Recall Action |
Recall |
Recall Action Instructions |
Replacement products made of more suitable material are being manufactured and will be provided to customers as soon as they become available. In the interim, customers are reminded on the correct usage of the device according to the surgical technique. |
Contact Information |
02 9417 9200 - Mathys Orthopaedics |