TGA Recall Reference RC-2013-RN-01017-1
Product Name/Description Xia 3 Titanium Torque Wrench & Xia 3 Titanium Torque Wrench Short (used for final tightening of the blocker on the head of the pedicle screw during posterior spinal fusion surgery)

Xia 3 Titanium Torque Wrench
Catalogue #: 48237028 (48230012-T, 48230004-T, 48230012NSW-T, 48230004NZ-T, WKSHP48230004-T)
Lot Numbers: 11E042, 11E047

Xia 3 Titanium Torque Wrench Short
Catalogue #: 482397028 (HANXIA3-T, HUNTSPINEEXT-T, 482370-T)
Lot Numbers: 11E036, R11E036, 11A957

ARTG Number: 160151
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 1/10/2013
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker Spine has received reports that the hex tip of the Xia 3 Torque Wrenches of an identified Lot Number subset have been fracturing during final tightening. The hex tip of the wrench is broken off and separated from the main tube.
Recall Action Recall
Recall Action Instructions Customers are requested to quarantine all affected lots. A Stryker representative will contact all affected customers to arrange for collection of the Xia 3 Torque Wrench. Unaffected replacement units will be provided.
Contact Information 02 9467 1214 - Stryker Regulatory Affairs and Quailty Assurance Office