Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00984-1
Product Name/Description HIV Ag/Ab Combo (CHIV) Assay (used with ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP) (an in vitro diagnostic medical device (IVD))

Catalogue Number: 06520528

Siemens Material Number: 10283020

Multiple lot numbers affected

ARTG Number: 205090
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 25/09/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens has determined that not all distributed lots would meet the performance criteria as currently described in the Instructions for Use (IFU) with respect to the analytical sensitivity. Siemens are updating the labelling to better reflect the observed and potential range of assay performance with respect to this parameter (i.e less than and equal to 2.0 IU/mL).
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is providing work around instructions for users to follow until the revised Instructions for Use (IFU) are released.
Contact Information 1800 310 300 - Siemens Technical Support Centre