Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00976-1
Product Name/Description Stryker Standard Video Cart and Vision Mounting Arm

Catalogue Numbers: 0240099011 (Video Cart), 0240099011I (Video Cart), 0240095200 (Mounting Arm)

All lots affected

ARTG Number: 106511
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 20/09/2013
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue If the tension screw present in the mounting yoke of the Vision Mounting Arm is loose and the user applies excessive force during positioning, the user can push the monitor out of the yoke. This will cause the monitor to dislodge from the mounting arm and fall potentially impacting an operator or patient.
Recall Action Recall for Product Correction
Recall Action Instructions Stryker is advising users that a new screw and washer will be installed on all affected units to ensure the monitor is securely fastened to the mounting yoke. It is expected that this will be completed by April 2014.
Contact Information 1800 803 601 - Stryker Australia