Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00936-1 |
Product Name/Description |
Howard Wright Limited Hospital Beds and Stretchers
Model Numbers: M7, M8 and M9
Manufactured between November 2011 and June 2013-09-16
ARTG Numbers : 134450, 160330 & 188353 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
6/09/2013 |
Responsible Entity |
|
Reason/Issue |
Howard Wright have conducted testing on a component of the siderail used in the M7, M8 and M9 model beds, and have found a variation in the force required to un-latch the siderail. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Howard Wright is providing update kits and a replacement procedure to be implemented at annual service by the customers maintenance provider. |
Contact Information |
1800 120 727 - Howard Wright Limited Customer Service |