Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00932-1
Product Name/Description Crocodile Size 3 Walker (walking aid)

Article Number: 86803
Recall Action Level Consumer
Recall Action Classification Class II
Recall Action Commencement Date 13/09/2013
Responsible Entity Mobility Plus Wheelchairs
Reason/Issue Mobility Plus is undertaking a recall for product correction following complaints of breakage of the bolt attaching the handle to the frame on Crocodile 3.
Recall Action Recall for Product Correction
Recall Action Instructions Mobility Plus is advising their customers and end users that replacement bolts are required to be fitted to the Crocodile Size 3 walker. Mobility Plus will arrange for the fitting of the new bolt. For more details, please see http://tga.gov.au/safety/alerts-device-crocodile-walker-130919.htm .
Contact Information 03 9495 1955 - Mobility Plus