Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00910-1
Product Name/Description DigitalDiagnost - EasyUpgrade DR (Diagnostic digital x-ray system)

system code 712086

ARTG Number: 117662
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/08/2013
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
The misinterpretation of the mirror icon to be a patient right side indication could potentially lead to a misdiagnosis.
Recall Action Recall for Product Correction
Recall Action Instructions Clinical staff should follow the Instructions for Use and use lead letters to indicate body orientation of the patient or equivalent electronic markers to provide Left/Right indicators, during all x-ray image acquisitions. Philips is advising users that a software and hardware upgrade will be performed to replace the mirror icon. Philips will also provide updated Instructions for Use for the work spot and the system.
Contact Information 1800 251 400 - Philips Customer Care Centre