Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00902-1
Product Name/Description Zimmer Trabecular Metal Reverse Glenosphere Distractor
(Supplied as part of the reusable instrumentation kit for shoulder replacement surgery)


Item number: 00-4309-049-00

Multiple lot numbers affected
ARTG Number: 109936
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 26/08/2013
Responsible Entity Zimmer Pty Ltd
Reason/Issue Zimmer is initiating a recall of the Trabecular Metal Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilised without proper lubrication. As a result, there is a potential for the device to fail to provide impact to the Glensophere when the trigger is pulled. In addition, there is a potential for the foot of the device to fracture during use which may result in the device being unable to remove the Glenosphere.
Recall Action Recall
Recall Action Instructions Zimmer is requesting users to remove affected lots from use. Zimmer are replacing affected stock with replacement product which incorporates a new design.
Contact Information 1800 803 263 - Zimmer Customer Service