Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00890-1
Product Name/Description Endowrist Instruments and Accessories for the da Vinci Surgical Systems (Used in endoscopic procedures)

Part Numbers of Affected Reprocessing Instructions: 550874-02 and 550882-05

ARTG Number: 132451
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/08/2013
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue Device Technologies Australia are implementing a revised Reprocessing Instructions for Endowrist instruments and accessories.
The changes include:
- To allow the use of mildly alkaline enzymatic cleaners, instead of only pH-neutral enzymatic cleaners.
- To clarify soaking prior to active scrubbing and rinsing in order to soften soil and make pre-cleaning easier.
- To augment scrubbing by spraying water through the ISI-supplier Luer nozzle at 2 bar pressure onto the tip of the instrument.
- To add magnification to the visual inspection of instruments to make it easier to ensure that instruments are free of visible soil prior to automated reprocessing.
Recall Action Recall for Product Correction
Recall Action Instructions Device Technologies Australia is providing on site training for the revised instructions, review updated content and provide a Cleaning and Sanitation kit (including the revised instructions) for all users.