Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00890-1 |
Product Name/Description |
Endowrist Instruments and Accessories for the da Vinci Surgical Systems (Used in endoscopic procedures)
Part Numbers of Affected Reprocessing Instructions: 550874-02 and 550882-05
ARTG Number: 132451 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
22/08/2013 |
Responsible Entity |
|
Reason/Issue |
Device Technologies Australia are implementing a revised Reprocessing Instructions for Endowrist instruments and accessories. The changes include: - To allow the use of mildly alkaline enzymatic cleaners, instead of only pH-neutral enzymatic cleaners. - To clarify soaking prior to active scrubbing and rinsing in order to soften soil and make pre-cleaning easier. - To augment scrubbing by spraying water through the ISI-supplier Luer nozzle at 2 bar pressure onto the tip of the instrument. - To add magnification to the visual inspection of instruments to make it easier to ensure that instruments are free of visible soil prior to automated reprocessing. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Device Technologies Australia is providing on site training for the revised instructions, review updated content and provide a Cleaning and Sanitation kit (including the revised instructions) for all users. |
Contact Information |
1800 429 551 - Device Technologies Australia |