Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00776-1 |
Product Name/Description |
Cell Marque CDX-2 (EPR2864Y*) antibody 7ml - an in vitro diagnostic medical device (IVD)
Catalogue Numbers: CM235R-17 and CM235R-18
Lot Number: 1221903 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
29/07/2013 |
Responsible Entity |
|
Reason/Issue |
Cell Marque CDX-2 is used to aid in the differentiation of colorectal adenocarcinoma and lung adenocarcinomas. The CDX-2 antibody typically stains the nuclei of colorectal carcinoma cells but not lung adenocarcinoma cells. Abacus has advised that the CDX-2 antibody may exhibit a light sporadic, focal, nuclear staining in a limited number of lung adenocarcinomas such as those with enteric differentiation, and may demonstrate slightly stronger staining than other lots. |
Recall Action |
Recall |
Recall Action Instructions |
Abacus ALS is requesting users to destroy affected stock. Replacement units will be provided as soon as possible. The requirement for a review of patient results reported from this lot should be determined by the laboratory director. |
Contact Information |
1800 287 287 - Abacus ALS Customer Service |