Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00714-1 |
Product Name/Description |
Fine-Medium Buchanan Pluggers (Dental instrument used during endodontic procedures)
Part Number: 952-0005
Lot Numbers: 4851810 & 4849060 Shipped from April to May 2013 ARTG Number: 176940 |
Recall Action Level |
Retail |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
16/07/2013 |
Responsible Entity |
|
Reason/Issue |
Ormco has received reports that the Buchanan Plugger .08 Taper, which should have a .08 laser mark on the shaft of the plugger, actually has a .06 laser mark instead. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Ormco are providing work around instructions for user to follow. |
Contact Information |
1800 052 303 - Ormco Customer Care |