Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00672-1
Product Name/Description ADVIA Centaur Homocysteine (HCY) Assay used with the ADVIA Centaur, Centaur XP, and Centaur CP systems

Multiple Catalogue numbers and Lot numbers affected

ARTG Number: 175075

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 3/07/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens have identified that the percent recovery for 1:10 diluted patient sample dilutions is lower than reported in the Instructions for Use (IFU). Dilution of patient samples is primarily used to resolve results above the diagnostic range of the HCY Assay.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising end users to ensure that 1:10 sample dilution testing is no longer carried out. The performance of 1:2 sample dilutions has been confirmed to meet the IFU recoveries and can continue to be tested. This action has been closed-out on 07/09/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre