Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00649-1
Product Name/Description Dimension Vista 500 and Dimension Vista 1500 using software versions 3.5.1 and 3.6 - Clinical chemistry analyser. An in vitro diagnostic medical device (IVD)

Siemens Material Numbers: 10488224 and 10444802

ARTG Number: 174699
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/06/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare has confirmed that, under specific rare conditions which require a system reset, there is a potential for a sample to be dispensed into an aliquot well that has previously been used for a QC or calibrator product from a Vista vial. If two samples are dispensed into the same aliquot well, the first sample will always be QC or calibrator from a Vista vial. Patient sample will never be dispensed into an aliquot well that contains another patient sample.
Recall Action Recall for Product Correction
Recall Action Instructions In the event of an instrument reset, end users are requested to check if there are any QC or calibration tests pending. If there are QC or calibrations pending a software restart is necessary before resuming patient processing. If there is no QC or calibration pending reset or troubleshoot the instrument as per the Operators Guide. A software update will be provided as a permanent fix.
Contact Information 1800 310 300 - Siemens Technical Support Centre