TGA Recall Reference RC-2013-RN-00625-1
Product Name/Description V60 Non-invasive Ventilators

Multiple serial numbers affected

ARTG Number: 134845
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/06/2013
Responsible Entity Mayo Healthcare Pty Ltd
Reason/Issue In the rare even a component fails due to a software issue, the following sequence of events may occur:

1) Power Management Board Assembly (PMBA) VMAIN shorts to ground,

2) the ventilator shuts down,

3) the software repeatedly cycles through part of its initialisation sequence and Power On Self Test (POST), the alarm may not sound to indicate that ventilation therapy has ceased.
Recall Action Recall for Product Correction
Recall Action Instructions The V60 ventilator may continue to be used in accordance with its directions for use. A Mayo healthcare representative will be contacting users to schedule the corrective action to upgrade the software. A loan unit will be provided to the users until the corrective action is completed.