TGA Recall Reference RC-2013-RN-00615-1
Product Name/Description Niti-S Biliary Covered Stent (Flare-Type) Nagi/Pseudocyst 10mm x 3cm

Model number: BS1003FW

Lot number: 13D1394D3EB780

Expiry date: 2016-04

ARTG number: 130531
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/06/2013
Responsible Entity Pyramed Pty Ltd
Reason/Issue The manufacturer has identified that there is a potential problem with the loosening of the Y connector on the hub of the delivery sheath which is outside the patient's body during delivery of the stent.The outer sheath and Y connector should be adjoined each other using glue as well as screw compression, however the gluing process was skipped in some cases with the affected lot number during manufacturing.
Recall Action Recall
Recall Action Instructions Pyramed is asking customers to return the affected stock for replacement with unaffected stock.
Contact Information 1800 651 882 - Pyramed