Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00589-1
Product Name/Description EZTest steam and Smart-Read EZTest steam Biological Indicators

Lot numbers: S-444, SR-444, S-445, SR-445, S-447

ARTG Number: 126974
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 14/06/2013
Responsible Entity Austmel Pty Ltd
Reason/Issue The manufacturer determined that the read out time is not within specification of the label claims on these lots. There is potential for delayed or incomplete colour change (false fail) at the indicated incubation time of 10 hours.
Recall Action Recall
Recall Action Instructions Austmel are asking customer to return all unused products for replacement with unaffected stock. If affected stock must be used, Austmel is advising end users to incubate the exposed biological indicator for a minimum of 72 hours.
Contact Information 1800 074 085 - Austmel Customer Service