Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00583-2
Product Name/Description Intuitive Surgical 5mm Cannula and 5mm Flared Cannula (Reusable endotherapy cannula)

5mm Cannula
Product code: INS-420011

5mm Flared Cannula
Product code: INS-420262

ARTG Number: 97289
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/06/2013
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue The manufacturer has received a small number of customer complaints regarding damaged 5mm cannula which could lead to difficulty inserting or utilising 5mm surgical instruments. Using a damaged or bent cannula can result in the loss of smooth instrument motion and/or particles of instrument shaft material being generated, which may drop into the patient.
Recall Action Recall for Product Correction
Recall Action Instructions Device Technologies is updating the current instructions for use with a modified inspection process for the identification of damaged or bent cannula using an obturator which will be provided to users.