Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00578-1
Product Name/Description Sonopet Console (Ultrasonic Surgical Aspirator System)

Catalogue number: 5450852000

Serial numbers: 1134000114, 1134000124, 1134000134, 1134000144, 1134000184, 1134000204 & 1209400174

ARTG number: 174364
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 6/06/2013
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has received reports that the Irrigation Pump may not function and audible alarm may not sound. In cases where the irrigation pump fails, saline would not be provided to the tip and the tip may overheat.
Recall Action Recall for Product Correction
Recall Action Instructions Stryker will be replacing affected units with a loan unit until the unit is repaired.
Contact Information 1800 803 601 - Stryker Regulatory Affairs and Quality Assurance Coordinator