Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00559-1 |
Product Name/Description |
SYNCHRON Systems Lactate (LACT). An in vitro diagnostic medical device (IVD).
Reference number: A95550
Lot number: M210077
Expiration date: 2013-10-31 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
3/06/2013 |
Responsible Entity |
|
Reason/Issue |
Beckman Coulter has confirmed customer reports of the affected lot failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions. |
Recall Action |
Recall |
Recall Action Instructions |
Beckman Coulter is requesting end users to discontinue use of the affected lot. Re-assessment of past results is not required as the LACT results produced after acceptable QC recoveries on each reagent cartridge are accurate. |
Contact Information |
1800 060 881 - Beckman Coulter Customer Support Centre |