Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00559-1
Product Name/Description SYNCHRON Systems Lactate (LACT). An in vitro diagnostic medical device (IVD).

Reference number: A95550

Lot number: M210077

Expiration date: 2013-10-31
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 3/06/2013
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue Beckman Coulter has confirmed customer reports of the affected lot failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions.
Recall Action Recall
Recall Action Instructions Beckman Coulter is requesting end users to discontinue use of the affected lot. Re-assessment of past results is not required as the LACT results produced after acceptable QC recoveries on each reagent cartridge are accurate.