Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00549-1
Product Name/Description Medrad Veris MR Vital Signs Monitor - Power cable

Catalogue number: 3010556
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 30/05/2013
Responsible Entity Imaxeon Pty Ltd
Reason/Issue Due to a latent design reliability issue there is potential for electrical shorting which can result in heating/melting of the cable jacket.
Recall Action Recall for Product Correction
Recall Action Instructions Imaxeon is providing additional safety instructions for end users until replacement DC power cables can be manufactured. Once completed the cables will be sent to end users along with instructions on returning the recalled cables.
Contact Information 02 8845 4999 - Imaxeon