Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00549-1 |
Product Name/Description |
Medrad Veris MR Vital Signs Monitor - Power cable
Catalogue number: 3010556 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
30/05/2013 |
Responsible Entity |
|
Reason/Issue |
Due to a latent design reliability issue there is potential for electrical shorting which can result in heating/melting of the cable jacket. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Imaxeon is providing additional safety instructions for end users until replacement DC power cables can be manufactured. Once completed the cables will be sent to end users along with instructions on returning the recalled cables. |
Contact Information |
02 8845 4999 - Imaxeon |