Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00536-1
Product Name/Description CAPD Solution Transfer Set with Locking Connector, Minicap Extended Life PD Transfer Set with Twist Clamp, Minicap Extended Life PD Transfer Set with Twist Clamp (Extra Short and Extra Long)

CAPD Solution Transfer Set with Locking Connector
Product code: 5C4160

Minicap Extended Life PD Transfer Set with Twist Clamp
Product code: 5C4482, R5C4482
Extra Short product code: 5C4483, R5C4483
Extra Long product code: R5C4484

ARTG Number: 167416
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/06/2013
Responsible Entity Baxter Healthcare Pty Ltd
Reason/Issue Baxter is providing additional information regarding correct device use and additional warnings to end users. The improper use of peritoneal dialysis (PD) transfer sets could result in contamination leading to infection or peritonitis. There is also the potential of thyroid dysfunction following iodine exposure in children and infants. The current IFU does not list this as a warning.
Recall Action Recall for Product Correction
Recall Action Instructions Baxter is providing additional information regarding correct device use and additional warnings to end users. This action has been closed-out on 01/02/2016.