Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00530-1 |
Product Name/Description |
MIA16 Minimally Invasive Attachment used with the Anspach
Product Insert version 18-0074 Rev A 05/12
ARTG Number: 141597 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
29/05/2013 |
Responsible Entity |
|
Reason/Issue |
During hydrogen peroxide sterilisation testing of the MIA16 device it was determined that the Sterrad 100S is not capable of achieving the expected sterility assurance level (SAL) of 10^-6. Therefore, Sterrad 100S should not be used for sterilising the MIA16 attachment. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
End users are advised to not use the Sterrad 100S to sterilise the MIA16 attachment. |
Contact Information |
1800 796 843 - Synthes Customer Service |