Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00522-1
Product Name/Description Intuitive Surgical EndoWrist Instrument Hot Shears Monopolar Curved Scissors

Part numbers: 420179-09 and -10, 400179-09 and-10
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 29/05/2013
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue Certain versions of the MSC instruments may develop micro-cracks near the distal (scissor) end of the shaft following reprocessing. The manufacturer identified these micro-cracks during internal testing of an unrelated design modification. These micro-cracks may create a pathway for electrosurgical energy to lead to patient tissue during use and potentially cause thermal injury.
Recall Action Recall
Recall Action Instructions Device Technologies is requesting their customers to follow work around instructions until alternative stock can be provided.
Contact Information 02 9975 5755 - Device Technologies