Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00522-1 |
Product Name/Description |
Intuitive Surgical EndoWrist Instrument Hot Shears Monopolar Curved Scissors
Part numbers: 420179-09 and -10, 400179-09 and-10 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
29/05/2013 |
Responsible Entity |
|
Reason/Issue |
Certain versions of the MSC instruments may develop micro-cracks near the distal (scissor) end of the shaft following reprocessing. The manufacturer identified these micro-cracks during internal testing of an unrelated design modification. These micro-cracks may create a pathway for electrosurgical energy to lead to patient tissue during use and potentially cause thermal injury. |
Recall Action |
Recall |
Recall Action Instructions |
Device Technologies is requesting their customers to follow work around instructions until alternative stock can be provided. |
Contact Information |
02 9975 5755 - Device Technologies |