Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00477-1
Product Name/Description Standard SlingBar 450

Distributed between January 1980 and November 2012

Product numbers: 20090029, 20190024, 20190041, 31190015, 31290017, 31290043, 3156001, 3156003, 3156007, 3156014, 3156016, 31590043 & 5019013

These product numbers may be bundled with the Golvo 1000, 3000 and 7000 series lifts, Viking M, Viking L, LikoLight 100 and 125, Caroli 101, and Uno 100, 101 and 102

ARTG Number: 188199
Recall Action Level Consumer
Recall Action Classification Class II
Recall Action Commencement Date 9/09/2013
Responsible Entity Hill-Rom Pty Ltd
Reason/Issue Hill-Rom has become aware of a safety issue related to the Standard SlingBar 450, which could be attached to various Liko overhead and mobile lifts. A serious injury to a patient, caregiver, or bystander could result from tripping or otherwise falling in to the slingbar, typically while the lift is parked or stored when not in use. The sling hook on the other end of the sling bar could pierce a body structure such as the head, shoulder, or groin.
Recall Action Recall for Product Correction
Recall Action Instructions Hill-Rom is providing customers with a sling-bar safety cover to mitigate the risk of patient or user injury. For more details, please see http://www.tga.gov.au/safety/alerts-device-standard-slingbar-450-131001.htm .
Contact Information 1800 147 118 - Hill-Rom Technical Support