Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00467-1 |
Product Name/Description |
Integra Ojemann Cortical Stimulator (Electrical antiseizure brain stimulator)
Catalogue number: OCS2
All serial numbers affected
ARTG number: 150717 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
17/05/2013 |
Responsible Entity |
|
Reason/Issue |
The manufacturer found that if the headphone jack is in use during a surgical procedure with the OCS2 and a non-intended voltage (such as static electricity) comes in contact with the outer case of the unit, that voltage could be transmitted to the patient and could be a potential source of injury. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Integra is advising users that the device may still be used providing the audio jack is not used. An Integra representative will contact affected customers to arrange for the disabling of the audio jacks and prevent the problem occurring. No reports of this issue have been received from the field. |
Contact Information |
1300 550 599 - Integra Customer Service |