Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00447-1
Product Name/Description Berichrom PAI (Plasmin activator inhibitors). An in vitro diagnostic medical device (IVD)

Catalogue number: OWOA15

Siemens material number: 10446642

Lot number: 541577
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 16/05/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue The incorrect Table of Assigned Values (TAV) was provided with the product. The use of the incorrect values for Standard 2 will result in slightly increase calibration curves, potentially resulting in decreased values for patient samples. The degree of deviation is less than 10 percent.
Recall Action Recall for Product Correction
Recall Action Instructions Customers who use the manual method to carry out the test are requested to use the assigned values provided in the recall correspondence. The values provided on the Lot Data CD for use with the BCS/BCS XP systems are correct.
Contact Information 1800 310 300 - Siemens Technical Support Centre