Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00422-1
Product Name/Description T4 Zipper Toga (XL - Tall) (Operating room gown)

Catalogue number: 0400850000

Lot numbers: Lots prior to 12040356

ARTG number: 141174
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/05/2013
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Testing by Stryker indicated that the sterilisation barriers were not valid for the stated 3 year shelf life and a breach of sterilisation risk is present for products older than 2 years.
Recall Action Recall
Recall Action Instructions Stryker is asking customers to perform inspection of product expiry dates and lot numbers and discard any affected products.
Contact Information 02 9467 1214 - Stryker Regulatory Affairs and Quality Assurance Coordinator