Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00420-1
Product Name/Description Soft Tissue Retractor, small, extendible

Part Number: 325.010

All Lot Numbers Affected
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/05/2013
Responsible Entity Synthes Australia Pty Ltd
Reason/Issue Synthes have received complaints about tip breakage and/or locking nut malfunction. There have been reports of retrieval of device fragments intraoperatively and delays in surgical procedures.
Recall Action Recall
Recall Action Instructions Customers are requested to return all affect units to DePuy Synthes Australia for replacement units.

Synthes will implement a design change to improve the robustness of material of construct of the locking nut.
Contact Information 1800 796 843 - Synthes Australia Customer Service