Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00397-1 |
Product Name/Description |
Brainlab Cranial Navigation System, all versions ARTG Number: 96517 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
3/05/2013 |
Responsible Entity |
|
Reason/Issue |
Brainlab has detected that, when using the Brainlab Cranial Navigation System, the following may have a significant effect on the overall navigation accuracy: - Large distance between reference array and region of interest - Major changes of the camera position relative to the reference array during the procedure |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Brainlab are providing additional instructions for use regarding the correct use of the Brainlab Cranial Navigation System. End users are advised to continue to follow the instructions in the user manual and adhere to the measures documented in the additional instructions for use (Measures to Improve Cranial Navigation Accuracy). |
Contact Information |
02 9424 3800 - Brainlab Australia |