Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00380-1 |
Product Name/Description |
A-dec 311 Chair and A-dec 361 Support Centre
A-dec 311 Chair ARTG Number: 118326
A-dec 361 Support Centre ARTG Number: 118333 Multiple Serial Numbers (see attached) |
Recall Action Level |
Retail |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
7/05/2013 |
Responsible Entity |
|
Reason/Issue |
The strength of some screws (or "fasteners") may have been compromised due to a plating process issue. Under certain circumstances, the fasteners can become brittle, which might lead to a chair and/or delivery system instability. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
The sponsor is advising their customers that the dealer will be replacing the fasteners with screws unaffected by hydrogen embrittlement. |
Contact Information |
02 8332 4000 - A-Dec Australia |