Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00378-1
Product Name/Description Tensioner (Instrument used for general orthopaedic repairs using implants like cerclage wires, cables and crimps/clamps)

Part Number: 03.311.001
Multiple lot numbers
ARTG Number: 203636
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 30/04/2013
Responsible Entity Synthes Australia Pty Ltd
Reason/Issue There have been reports of the instrument falling apart and/or the K-Wire being trapped inside. The connection between the reaction fork and the main body is connected without a gap and secured with tightening of 23Nm. There may be instruments which have not been tighten with adequate torque.
Recall Action Recall for Product Correction
Recall Action Instructions Synthes is requesting that customers return the affected devices for reworking by the manufacturer.
Contact Information 1800 796 8437 - Synthes Australia Customer Service