Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00371-1
Product Name/Description RPM Respiratory Gating System, version v1.7.5 with 3D Option (Patient respiratory monitoring system used with radiation therapy systems)

Product code: H50

Multiple serial numbers affected

ARTG Number: 117951
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 24/04/2013
Responsible Entity Varian Medical Systems Australasia Pty Ltd
Reason/Issue While operating with the 6-dot marker block, this version of RPM Gating may produce an inverted respiratory waveform in amplitude and phase-gating sessions. The error may be reproduced if the RPM System is induced to frequently lose and reacquire tracking of the 6-dot marker block during a calibration sequence, a reference session or a single session after tracking has begun.
Recall Action Recall for Product Correction
Recall Action Instructions Varian is providing temporary work around instructions and is implementing a software update to permanently correct the problem.
Contact Information 1800 657 036 - Varian Oncology Helpdesk