Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00367-1
Product Name/Description VITROS Immunodiagnostic Products CK-MB Calibrators. An in vitro diagnostic medical device (IVD)

Product Code: 1286293
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/04/2013
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue Ortho Clinical Diagnostics has revised the storage instructions for VITROS Immunodiagnostic Products CK-MB Calibrators. Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20degC generated invalid calibrations that were detected by positively biased quality control results.
Recall Action Recall for Product Correction
Recall Action Instructions OCD is updating the instructions for use to reflect that refrigerated storage at 2-8 degrees centigrade is acceptable. OCD is asking users to discontinue use of units stored at -20 degrees centigrade. This action has been closed-out on 11/02/2016.
Contact Information 1800 032 0359 - Ortho-Clinical Diagnostics Customer Technical Services