Product Name/Description SYNCHRON Systems Lactate (LACT), An in vitro diagnostic medical device (IVD)

Reference: A95550

Lot Number: M206209

Expiry: 2013-04-30
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/04/2013
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue Beckman Coulter has received customer reports of suppressed LACT results for samples containing LACT concentration of 6.5 mmol/L or greater, when using the affected lot number. Internal testing confirmed that the issue for samples containing LACT concentrations between 9.2 and 11.0 mmol/L. Although these results are within the analytical range of 0.3 - 11.0 mmol/L, the results are suppressed and the system identifies the error condition RXN RATE HL.
Recall Action Recall
Recall Action Instructions Beckman Coulter is asking users to cease using the affected lot and is offering a replacement.