Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00358-1
Product Name/Description Panbio Dengue IgM Capture ELISA, An in vitro diagnostic medical device (IVD)

Catalogue numbers: E-DEN01M / E-DEN01M05

Lots: 12306, 12348 & 13007

ARTG Number: 22341
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/04/2013
Responsible Entity Inverness Medical Innovations Australia Pty Ltd T/A Alere
Reason/Issue Alere have received complaints of invalid test runs due to results falling outside the cut-off value customer acceptance range leading to an invalid test. Analysis of retained kits and customer results has indicated that the invalid test runs are due to the customer acceptance range being inappropriately assigned for these batches. The effect of the inappropriate acceptance range is that a greater number than normal of test results will be considered invalid, requiring the test to be repeated.
Recall Action Recall for Product Correction
Recall Action Instructions Alere is providing updated reference values for the affected kits.
Contact Information 07 3363 7711 - Alere Technical Support