Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00355-1
Product Name/Description Liquichek ToRCH Plus IgM Control - Positive (An in vitro diagnostic medical device (IVD))

Catalogue Numbers: 229 and 229PX
Lot Number: 18972
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 19/04/2013
Responsible Entity Bio-Rad Laboratories Pty Ltd
Reason/Issue This product contains detectable levels of HCV RNA which originated from the non-specific human IgM component used in the mouse monoclonal antibody conjugation process for this product.
Recall Action Recall
Recall Action Instructions Bio-Rad is asking users to destroy the affected stock using appropriate disposal methods.
Contact Information 1800 224 354 - Bio-Rad Customer Service