Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00346-1
Product Name/Description Optima MR430s and MSK 1.5T Extreme (MRI System)

Scanner identification numbers between '06 01 2008-001' and '01 13 2013-001' (format "mm dd yyyy-nnn")

ARTG Number: 193173
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 18/04/2013
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue The GE Extremity 1.5T MR Scanners use a liquid coolant to protect the gradient coil inside the magnet from heating. If the coolant supply lines are assembled incorrectly, they may leak the liquid onto the floor which may result in a slippery surface and fall hazard.
Recall Action Recall for Product Correction
Recall Action Instructions GE is correcting the affected units and is asking users to monitor for leakage until the units are corrected.
Contact Information 1300 722 229 - GE National Service and Support Centre