Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00334-1
Product Name/Description Elekta Integrity Digital Linear Accelerator, version 1.1

ARTG Number: 191025
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/04/2013
Responsible Entity Elekta Pty Ltd
Reason/Issue If the "static tolerances" in the calibration files (loaded database) for the unwanted movement during the beam delivery are incorrect, the unwanted movement may not terminate treatment. This is only a problem if a malfunction such as a brake failure or an equivalent fault occurs during beam delivery.
Recall Action Recall for Product Correction
Recall Action Instructions Elekta is advising that the solution to the problem is to apply the Integrity 1.1 Static Tolerance Updater Tool. This tool will need to be re-applied, if a new Linac license is applied in the future. A complete solution will be included in the next release of the software.
Contact Information 02 8907 1800 - Elekta Regulatory Affairs