Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00331-1 |
Product Name/Description |
Hill-Rom VersaCare VC 755 Bed
Model: P3201K
ARTG Number: 199714 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
6/05/2013 |
Responsible Entity |
|
Reason/Issue |
Due to high mains input supply power the Power Control PC board may fail resulting in loss of function. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Hill-Rom is developing a correction to the Power Control PC board to compensate for the higher mains power. A Hill-Rom technician will schedule a visit to their customers facility to replace the PC board of affected beds. |
Contact Information |
1800 147 118 - Hill-Rom Technical Support |