Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00331-1
Product Name/Description Hill-Rom VersaCare VC 755 Bed

Model: P3201K

ARTG Number: 199714
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/05/2013
Responsible Entity Hill-Rom Pty Ltd
Reason/Issue Due to high mains input supply power the Power Control PC board may fail resulting in loss of function.
Recall Action Recall for Product Correction
Recall Action Instructions Hill-Rom is developing a correction to the Power Control PC board to compensate for the higher mains power. A Hill-Rom technician will schedule a visit to their customers facility to replace the PC board of affected beds.
Contact Information 1800 147 118 - Hill-Rom Technical Support