Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00320-1
Product Name/Description EVOTECH Endoscope Cleaner and Reprocessor (ECR) System (Endoscope disinfecting unit)

Product code: 50004

All lot/serial numbers affected.

ARTG number: 146438
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 12/04/2013
Responsible Entity Johnson & Johnson Medical Pty Ltd
Reason/Issue The manufacturer has determined that under certain circumstances the EVOTECH Endoscope Cleaner and Reprocessor System may not properly calculate the expiration date of the CIDEZYME XTRA Multi Enzymatic Detergent from the information entered by the user. In these circumstances, the expiration date displayed on the EVOTECH ECR screen is one month after the labelled expiration date.
Recall Action Recall for Product Correction
Recall Action Instructions Johnson & Johnson is providing instructions on how to identify if there is any discrepancy between the calculated expiration date and the labelled expiration date. Users are advised to adhere to the labelled expiration date.
Contact Information 1800 252 194 - Johnson & Johnson Medical Customer Service