Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00319-1
Product Name/Description EVOTECH Endoscope Cleaner and Reprocessor (ECR) System (Endoscope disinfecting unit)

Product number: 50004
ARTG: 146438
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/04/2013
Responsible Entity Johnson & Johnson Medical Pty Ltd
Reason/Issue Johnson & Johnson Medical is issuing a product correction for the EVOTECH Endoscope Cleaner and Reprocessor system after receiving reports of smoke emanating from the system and upon further investigation has discovered that a circuit board malfunction may occur in the Channel Control Manifold sub-assembly. This failure can potentially cause the following outcomes for the cycles:
1. Full Reprocessing Cycle (Cleaning and HLD) - If this issue were to occur while running a full automated cleaning and high level disinfection (HLD) cycle, there is potential for residual rinse water and/or alcohol to remain in the endoscope channels after the cycle completes,
2. HLD-Only Cycle - If this issue were to occur during reprocessing of endoscopes qualified for the HLD only cycle, there is potential for an inadequate amount of high-level disinfectant solution to be sent through some endoscope channels.
Recall Action Recall for Product Correction
Recall Action Instructions Johnson & Johnson Medical Pty Ltd is providing the users with the instructions to prevent the issues from occurring.
Contact Information 1800 252 194 - Johnson and Johnson Medical Customer Service