Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00318-1
Product Name/Description ADVIA Chemistry Systems Liquid Specific Protein Calibrator (an in vitro diagnostic medical device (IVD))

Siemens Material Number (SMN): 10311929

Part Number: B03-4845-01

Lot Number: 244560

Expiration Date: 2014-05-28

ARTG Number: 185171
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 15/04/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue The System Specific Values (SSVs) for the Rheumatoid Factor (RF) assay have been reassigned for this lot. Siemens has determined that the Rheumatoid Factor SSVs currently assigned to this lot may cause an average -15% shift in quality control and sample recovery.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is providing updated system specific values for the affected lot as a work around solution.
Contact Information 1800 310 300 - Siemens Technical Support Centre