Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-00318-1 |
Product Name/Description |
ADVIA Chemistry Systems Liquid Specific Protein Calibrator (an in vitro diagnostic medical device (IVD))
Siemens Material Number (SMN): 10311929
Part Number: B03-4845-01
Lot Number: 244560
Expiration Date: 2014-05-28
ARTG Number: 185171 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
15/04/2013 |
Responsible Entity |
|
Reason/Issue |
The System Specific Values (SSVs) for the Rheumatoid Factor (RF) assay have been reassigned for this lot. Siemens has determined that the Rheumatoid Factor SSVs currently assigned to this lot may cause an average -15% shift in quality control and sample recovery. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is providing updated system specific values for the affected lot as a work around solution. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |