Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00317-1
Product Name/Description IMPAX 5.2 DB Server, version 5.2 (Picture archiving and communication system, PACS)

ARTG Number: 127983
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/04/2013
Responsible Entity AGFA HealthCare Australia
Reason/Issue System downtime and/or slow performance may result if software in use is not at supported levels. To be specific, it has been identified that Oracle version 10.1.0.4.0 or older in use with IMPAX 5.2 DB Server can cause Oracle blocking sessions due to defect 360595 with fix available in Oracle 10.2.0.2 and subsequent versions.
Recall Action Recall for Product Correction
Recall Action Instructions Agfa HealthCare is advising that the ‘End of Service Life’ for IMPAX 5.2 was December 31, 2010. In order for Agfa HealthCare to continue to provide support, it is critical that sites upgrade to the latest version of IMPAX. In the interim, Agfa HealthCare strongly recommends installing the latest Oracle critical patch upgrades. Please see Oracle’s website for more details: http://www.oracle.com/technetwork/database/windows/downloads/index-101290.html .
Contact Information 1300 364 612 - Agfa Healthcare Service Help Desk