TGA Recall Reference RC-2013-RN-00301-1
Product Name/Description Stryker Suction Irrigation Tips (general purpose irrigation/aspiration units)

Multiple product codes and lot numbers
ARTG Number: 142830
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 4/04/2013
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue The pre-vacuum steam sterilisation temperature in the Instructions for Use (IFU Part # 1000-400-100 Rev. K) is incorrectly listed as 121 - 123 degrees centigrade, instead of the correct 132 - 133 degrees centigrade (IFU Part # 1000-400-100 Rev. L).
Recall Action Recall for Product Correction
Recall Action Instructions Stryker is correcting the Instructions For Use.
Contact Information 02 9467 1214 - Stryker Regulatory Affairs