Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00292-1
Product Name/Description PCA LifeCare Infusion System

Catalogue number: 0F7822703

Number on pump: 0F782

ARTG number: 138109
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 19/04/2013
Responsible Entity Hospira Pty Limited
Reason/Issue Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in serious injury. This issue is caused by wear and tear on the Half Nut which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
Recall Action Recall
Recall Action Instructions Hospira is recalling the affected pumps. The pumps have not been supplied since 2006.
Contact Information 1300 046 774 (Option 1) - Hospira Customer Service