Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00281-1
Product Name/Description CARDIOHELP-1, software version 3.3.0 (Heart-lung bypass system)

Serial Number: All devices up to 90410628 excluding S/N 90410622 and 90410626

ARTG Number: 179014
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 28/03/2013
Responsible Entity Maquet Australia Pty Ltd
Reason/Issue 1 - There have been reported cases of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touchscreen) of the CARDIOHELP device. All manual settings are lost and the system will revert back to the original clinical settings.
2 - Through internal testing, it has been determined that there is a possibility that electromagnetic interference due to electric current fluctuations may cause pressure valves on the display to "freeze".
Recall Action Recall for Product Correction
Recall Action Instructions Maquet is providing work around instructions to prevent risk of injury to the patient in the unlikely event of the failure mode occurring during use. Maquet has updated its current CARDIOHELP-i software to version 3.3.2.0 and this will be installed by an authorised Maquet service technician.
Contact Information 1300 304 372 - Maquet Australia