Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00277-1
Product Name/Description AU5800 Clinical Chemistry Analyser; all instrument serial numbers with software version 4.11 and prior (An in vitro diagnostic medical device (IVD))

ARTG Number: 177999
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/03/2013
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue A software error has been identified that has the potential to cause the reporting of erroneous patient results due to mis-identification by the AU5800 system.

This issue only occurs if all the following conditions apply:
a) the system is in "MEASURE 1" mode during analysis (Sequential, Barcode or Rack number mode)
b) a manual batch requisition is initiated
c) LIS is running in real time for TRI (TEST REQUESTING INFORMATION) Receive
d) when events b and c (above) occur simultaneously
Recall Action Recall for Product Correction
Recall Action Instructions End users are advised of work around procedures to ensure erroneous results are not reported. Beckman Coulter will be addressing this issue in a future software update.
Contact Information 1800 060 881 - Beckman Coulter Customer Support Centre