Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00275-1
Product Name/Description AQURE POC Management System, software versions 1.4.0 to 1.5.3 (An in vitro diagnostic medical device (IVD))
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/03/2013
Responsible Entity Radiometer Pacific Pty Ltd
Reason/Issue Patient results are displayed with the incorrect unit of measure in AQURE when:
a) AQURE is changed from the default setting for any or all parameters of International System of Units to use Non-International System of Units (SI Units); and
b) The result is outside reference and critical ranges; and
c) AQURE is being used to view the results.
Results on the analyser and reported to the HIS/LIS remain correct.
Recall Action Recall for Product Correction
Recall Action Instructions End users are advised to view patient results in their HIS/LIS system and not in AQURE POC management system. Radiometer will provide an update to software version1.5.4 to correct this issue.
Contact Information 1800 247 254 - Radiometer Pacific