Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00270-1
Product Name/Description Oxoid Antimicrobial Susceptibility Testing discs for Ampicillin (AMP10). An in vitro diagnostic medical device (IVD).

Product Code: CT0003B
Lot Number: 1215912
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/03/2013
Responsible Entity Thermo Fisher Scientific Australia & New Zealand
Reason/Issue Investigations have confirmed that cartridges within this batch may contain individual discs with insufficient amount of antibiotic. Continued use of this lot could result in false indication of resistance to ampicillin.
Recall Action Recall
Recall Action Instructions End users are requested to inspect their stock for affected units. Affected units are to be removed from further use. The requirement for the review of patient results should be determined by the laboratory director.
Contact Information 08 8238 9044 Oxoid Regulatory Affairs