Type of Product Medical Device
TGA Recall Reference RC-2013-RN-00255-1
Product Name/Description Elekta PreciseTreatment Table (Intended as a universal patient treatment support and positioning table for radiation therapy and simulation use)

Ranges: 124001 to 126924, 133801 to 133999, 213000 to 213189

ARTG Number: 165045
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 22/03/2013
Responsible Entity Elekta Pty Ltd
Reason/Issue The Column Rotation can accidentally be moved to a different position while in the locked position. This can occur between fields and can cause clinical mistreatment.
Recall Action Recall for Product Correction
Recall Action Instructions Elekta has provided work around instructions to ensure the treatment table is in the correct position. If end users determine that it is possible to rotate the column with the brakes applied they must contact their service representative. Elekta is currently developing a correction to repair the problem.
Contact Information 02 8907 1800 - Elekta Regulatory Affairs